Outputting Profile Data
Analog waveforms of the data on height of 126 lines are output.
Note:
Example: Z500-SW17
Refer to page 181 for details on output timing.
CHECK
The setting can be performed through the adjustment menu.
CHECK
Refer to page 28.
Item
Selection
OUT0 to OUT7
Output
Profile
(OUT0*)
Select the output range of profile data.
F.S.*
Profile
1/2F.S.
(See note.)
1/4F.S.
The range of profile varies depending on the Sensor connected.
Selection
Output range
F.S.
-20mm to +20mm
1/2F.S.
-10mm to +10mm
1/4F.S.
-5mm to +5mm
REFERENCE
Menus for Expert Menu
Description
Select Profile.
The output covers the whole measurement range.
The output covers one-half of the measurement
range.
The output covers one-quarter of the measurement
range.
The asterisk (*) indicates the default setting.