SECTION 4
Menus for Expert Menu
#2
Set Calibration
Set calibrations to compensate the measurement values. Set calibrations to compensate for differences
between measurement values and actual dimensions due to the color or material of the measurement
workpiece. The setting method varies depending on the measurement items set for the output number.
CHECK
CHECK
Compensate the Measured Height and Position
Workpiece height: 10 mm
Use the method shown below to calibrate workpieces so that a displacement value of 9.95 mm is
measured as 10 mm.
Measurement of point 1
Measure without
the workpiece.
Z500
78
Operation Manual
Compensation
Measurement items
Compensate the
Height, Edge position, Edge
measured height
center, Peak/bottom
and position.
Compensate the
measured level
Step: 2 pts, Step: 3 pts, Width
difference and
width.
Output can be used for fine adjustment after calibration setting. Use Output also to
compensate the measurement values when Define has been selected for the
measurement item.
REFERENCE
Refer to page 84.
Refer to page 83 for information on how to clear calibration settings.
To compensate for the error, perform the measurement twice.
Applicable measurement items: Height, Edge position, Edge center,
Peak/bottom
Example 1: When selecting Height
10mm
9.95mm
Calibrate to 10 mm.
Input the height
of point 1.
0mm
For setting, repeat the measurement
twice.
REFERENCE
For setting, repeat the measurement
once.
REFERENCE
Measurement value
Compensate for this error.
Displacement of workpiece
10mm
Measurement of point 2
Measure with the
workpiece in position.
Description
Refer to page 78.
Refer to page 81.
Input the height
of point 2.
10mm