Omron Z500 Operation Manual page 105

Profile measuring system
Hide thumbs Also See for Z500:
Table of Contents

Advertisement

Terminal Block Output
Judgment results can be output via the terminal block. The settings are enabled
for all output numbers.
Note:
For details on I/O Format, refer to page 154.
CHECK
The setting can be performed through the adjustment menu.
CHECK
Refer to page 28.
Selection/
Item
Setting range
Judge output A*
Format
Judge output B
0 to 99999
OFF delay
(0*)
GATE ON delay
1.0 to 1000.0ms
(See note.)
(1.0ms*)
GATE ON time
1.0 to 1000.0ms
(See note.)
(5.0ms*)
The GATE signal is output only when the Trigger is set to External or Free.
If the Trigger is not set, the GATE signal will always stay OFF regardless of the
settings of the GATE ON delay and GATE ON time.
REFERENCE
Menus for Expert Menu
Description
Select this function to output the details of the
judgment (HIGH/PASS/LOW/ERROR).
The output includes the judgment results for
OUT0 through OUT3. Those for OUT4 through
OUT7 will not be output.
Judgment results for OUT0 through OUT7 will be
output in the form of OK or NG.
OK : PASS
NG : HIGH, LOW, ERROR
Set the OFF delay time N (sampling time
Set a time that allows the external device to input
the result.
Set the time from when the result is output to the
terminals to when the GATE signal is turned ON.
The time to wait until stable output data can be
obtained. Set this time longer than the data t
ransmission delay time to external devices.
Set the length of time that the GATE signal
remains ON. Set a value so that the external
device can obtain the measurement result.
The asterisk (*) indicates the default setting.
SECTION 4
×
N).
Z500
103
Operation Manual

Advertisement

Table of Contents

Troubleshooting

loading

This manual is also suitable for:

Z500-mc15eZ500-mc10e

Table of Contents