How to exit from the E5061A/E5062A Measurement View
You need to exit from the E5061A/E5062A Measurement View to perform some
troubleshooting. The following is the procedure to exit from the E5061A/E5062A
Measurement View.
Step 1. Connect the mouse and external keyboard to the connectors on the E5061A/E5062A rear
panel.
Step 2. Turn the instrument on.
Step 3. Press
Step 4. Click
4-1.
Figure 4-1
Password dialog box
Step 5. Enter the password
Step 6. Click
View, then.windows desktop screen appears with 4 icons (My Computer, Recycle Bin, My
Network Places, and Network Analyzer.)
NOTE
If you wish to return to the Measurement View, double-click "Network Analyzer" icon.
NOTE
If you need to shut down the E5061A/E5062A and again turn on, perform in accordance
with the following procedure.
a. To get "Start" menu bar displayed, move the pointer to the bottom of the screen with
mouse.
b. Click "Start" and "Shut Down..." in the pull down menu. "Shut Down Windows" dialog
box opens.
c. Select "Shut down" button in the pull down menu.
d. Click "OK" button in the dialog box.
Chapter 4
How to exit from the E5061A/E5062A Measurement View
key.
Service Menu
-
Service Function
in the Password box and then click "OK" button.
kid
in Service Functions Menu. Then the E5061A/E5062A exit the Measurement
Exit
. Password dialog box opens as shown in Figure
Troubleshooting
89