How To Exit From The E5071C Measurement View - Agilent Technologies E5071C Service Manual

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

How to exit from the E5071C Measurement View

You need to exit from the E5071C Measurement View to perform some troubleshooting.
The following is the procedure to exit from the E5071C Measurement View.
Step 1. Connect the mouse and external keyboard to the connectors on the E5071C rear panel.
Step 2. Turn the instrument on.
Step 3. Press
Step 4. Click
4-1.
Figure 4-1
Password dialog box
e5071cse1057
Step 5. Enter the password
Step 6. Click
then.windows desktop screen appears with 3 icons (Network Analyzer, Recycle Bin, and
Internet Explorer.)
NOTE
If you wish to return to the Measurement View, double-click "Network Analyzer" icon.
NOTE
If you need to shut down the E5071C and again turn on, perform in accordance with the
following procedure.
a. To get "Start" menu bar displayed, move the pointer to the bottom of the screen with
mouse.
b. Click "Start" and "Turn off Computer" in the pull down menu. "Turn off Computer"
dialog box opens.
c. Click "Turn off" button in the dialog box.
Chapter 4
How to exit from the E5071C Measurement View
key.
-
Service Menu
Service Function
in the Password box and then click "OK" button.
kid
in Service Functions Menu. Then the E5071C exit the Measurement View,
Exit
. Password dialog box opens as shown in Figure
Troubleshooting
177

Advertisement

Table of Contents
loading

Table of Contents