System Dynamic Range Test; Description; Test Equipment - Agilent Technologies E5071C Service Manual

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Performance Test

6. SYSTEM DYNAMIC RANGE TEST

6. SYSTEM DYNAMIC RANGE TEST

Description

This test checks the system dynamic range for the receiver ports of the E5071C. The
system dynamic range is tested by performing an "Isolation" measurement 16 times with
segment sweep points for specified frequency ranges (after the response and isolation
calibrations are performed) and calculating the RMS deviation value from the 16
measurement data for each sweep frequency point. The maximum RMS deviation value in
each frequency range is extracted to determine the system dynamic range performance.

Test equipment

Coaxial cable with Type-N (m) connectors,
61 cm (24 in)
Type-N(m) 50 Ω Load termination (2 ea.)
36
Agilent N6314A(p/n 8120-8862)
part of Agilent 85032F/54D/33E/50D
with Type-N adapter