Making Mixer Measurements (Option 089 Only)
High Dynamic Range Swept RF/IF Conversion Loss

Perform the High Dynamic Range Measurement

1. Return the analyzer to the IF frequency range. Press
Start
100
M/µ
Stop
1
.
G/n
2. Make the connections shown in
Figure 2-19
3. Set the LO source to the desired CW frequency of 1500 MHz and power level to 13 dBm.