Adjusting The Circular Level Of The Instrument And Tribrach - Leica TS03 User Manual

Table of Contents

Advertisement

Messages
13.6
Adjust the circular
level step-by-step
180
6.
Either:
Press More to measure another set to the same target point.
The final adjustment values will be the calculated average from
all the measurements.
Press Cont to save the new adjustment data, or
Press ESC to exit without saving the new adjustment data.
The same messages or warning as in "13.3 Adjusting Line-of-Sight and Vertical
Index Error" may appear.

Adjusting the Circular Level of the Instrument and Tribrach

3
0016350_001
1.
Place and secure the tribrach onto the tripod, and then secure the
instrument onto the tribrach.
2.
Using the tribrach footscrews, level the instrument with the elec-
tronic level. To activate the electronic level, turn on the instrument,
and, if tilt correction is set to On, the Level & Plummet screen
appears automatically. Alternatively, press the
any app and select Level.
3.
The bubbles of the instrument and tribrach levels must be centred. If
one or both circular levels are not centred, adjust as follows.
Instrument: If the bubble extends beyond the circle, use the Allen
key supplied to centre it with the adjustment screws.
Tribrach: If the bubble extends beyond the circle, adjust it using the
adjustment pin in conjunction with the adjustment screws. Turn the
adjustment screws:
To the left: and the bubble approaches the screw.
To the right: and the bubble goes away from the screw.
4.
Repeat step 3. on the instrument and tribrach until both circular lev-
els are centred and no further adjustments are necessary.
After the adjustment, no adjustment screw should be loose.
3
1
key from within
Check & Adjust

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ts07

Table of Contents