Check and adjust
step-by-step
Messages
Check & Adjust
1.
Level the instrument with the electronic level. Refer to "4 Operation"-
"Level up with the electronic level step-by-step".
2.
0016346_001
3.
Press Store to measure to the target point.
4.
180°
0016347_001
☞
For checking the horizontal aim, the difference in Hz and V
are displayed.
5.
Press Store to measure to the target point.
☞
The old and new calculated values are displayed.
6.
Either:
Press More to measure another set to the same target point.
•
The final adjustment values will be the calculated average from
all the measurements.
Press Cont to save the new adjustment data, or
•
Press ESC to exit without saving the new adjustment data.
•
The following are important messages or warnings that may appear.
Messages
V-Angle is not suitable
for adjustment or wrong
face!
Out of Tolerance!
Previous values retained!
Aim at a point approximately 100
m from the instrument which is
within 5° of the horizontal.
± 9°
Change face and aim at the tar-
get point again
180°
Description
The vertical angle deviates from the required
horizontal / line-of-sight. Aim correctly at the
target point. Confirmation of the message
required.
Computed values out of tolerance. The previ-
ous values are retained and measurements
should be repeated. Confirmation of the
message required.
177