Check and adjust
step-by-step
Messages
Check & Adjust
1.
Level the instrument with the electronic level. Refer to "4 Operation"-
"Level up with the electronic level step-by-step".
2.
0016346_001
3.
Press Store to measure to the target point.
4.
180°
0016347_001
For checking the horizontal aim, the difference in Hz and V
are displayed.
5.
Press Store to measure to the target point.
The old and new calculated values are displayed.
6.
Either:
Press More to measure another set to the same target point.
The final adjustment values will be the calculated average from
all the measurements.
Press Cont to save the new adjustment data, or
Press ESC to exit without saving the new adjustment data.
The following are important messages or warnings that may appear.
Messages
V-Angle is not suitable
for adjustment or wrong
face!
Out of Tolerance!
Previous values retained!
Aim at a point approximately 100
m from the instrument which is
within 5° of the horizontal.
± 9°
Change face and aim at the tar-
get point again
180°
Description
The vertical angle deviates from the required
horizontal / line-of-sight. Aim correctly at the
target point. Confirmation of the message
required.
Computed values out of tolerance. The previ-
ous values are retained and measurements
should be repeated. Confirmation of the
message required.