Adjusting The Circular Level Of The Instrument And Tribrach - Leica FlexLine TS02 User Manual

Hide thumbs Also See for FlexLine TS02:
Table of Contents

Advertisement

11.5
Adjust the circular
level step-by-step
Check & Adjust

Adjusting the Circular Level of the Instrument and Tribrach

3
TSOX_067
1. Place and secure the tribrach onto the tripod, and then secure the instrument
onto the tribrach.
2. Using the tribrach footscrews, level the instrument with the electronic level. To
activate the electronic level, turn on the instrument, and, if tilt correction is set
to 1- or 2-axis, the Level/Plummet screen appears automatically. Alternatively,
press FNC from within any application and select Level/Plummet.
3. The bubbles of the instrument and tribrach levels must be centered. If one or
both circular levels are not centered, adjust as follows.
Instrument: If the bubble extends beyond the circle, use the Allen key supplied
to center it with the adjustment screws.
Tribrach: If the bubble extends beyond the circle, adjust it using the adjustment
pin in conjunction with the adjustment screws. Turn the adjustment screws:
• To the left: and the bubble approaches the screw.
1
2
3
FlexLine, 245