Performance Test

Set Dose
Acceptable O
Value
2
Acceptable NO
Value
2
Acceptable NO Value
Part No. 20568 Rev-02
2015-08
2. Turn the integrated backup INOMAX delivery to
3. Allow monitored values to stabilize (may take
40 ppm
95% ± 3 %
< 1.5 ppm
35-45 ppm
ON (250 mL/min.).
Verify "Backup ON" alarm occurs.
up to 3 minutes).
Verify the NO and NO
2
following ranges:
NO = 14-26 ppm
4.
Turn the backup INOMAX delivery OFF.
1. Using the pre-use set-up connectors, verify that
the O
flowmeter is set to 10 L/min.
2
2. Press NEXT button to automatically set the
INOMAX dose to 40 ppm.
3. Allow monitored values to stabilize (may take
up to 3 minutes).
Verify the NO, NO
and FiO
2
within the ranges in the performance test table.
4. Performance test is complete.
Press NEXT button to set the INOMAX dose to
zero.
If a monitored value is outside the range
Note:
indicated, see Section 4/ Alarms and
Troubleshooting Help.
readings are within the
NO
≤ 1.0 ppm
2
readings are
2
2-9