Alarm Help

• If an alarm occurs, safeguard the patient first before troubleshooting or repair
WARNING:
procedures.
• Use caution when troubleshooting the INOmax DS
patient.
• When possible, replace the unit in question and perform troubleshooting
procedure once the unit is removed from the MR scanner room.
If the system fails to operate properly:
1. Check the patient condition and take appropriate action.
2. Use the INOblender (see INOblender Operation Manual) or backup if necessary.
3. Verify that the system is set up as detailed in 3/ Patient Application.
4. Find a symptom or alarm condition in the troubleshooting table which best describes the problem and
follow the recommended actions to resolve the problem.
If the problem cannot be corrected:
Contact the Authorized Representative listed on the back cover of the operation manual.
If the INOmax DS
Plus MRI must be returned for servicing:
IR
Contact the Authorized Representative listed on the back cover of the operation manual.
4-12
Plus MRI while in use for a
IR
To activate on-screen alarm help, press the
Alarm Help Button next to the Alarm Silence
button.
Part No. 20568 Rev-02
2015-08