Inomax Cylinder 88-Size Duration Chart - Mallinckrodt INOmax DSIR Pocket Manual

Delivery system, series 3 software
Hide thumbs Also See for INOmax DSIR:
Table of Contents

Advertisement

INOMAX Cylinder 88-Size
For an 88-Size 800 ppm Cylinder Concentration*
(Illustrative Only)
5
10
20
40
80
This chart is representative of a range of doses available
on the INOmax DS
intended as the recommended therapeutic dose.
* All calculations in the table above are based on a full cylinder, 138 bar (2000
psig), 1963 liter "88" cylinder, with a cylinder change at 14 bar (200 psig). The
figures are calculated based on a total continuous breathing circuit gas flow and
a cylinder conversion factor of 14.2 liters per bar/0.98 liters per psig.
• INOMAX flow = [Desired dose × total ventilator flow] ÷ [Cylinder concentration
- desired dose]
• Cylinder volume = Cylinder conversion factor × cylinder pressure (bar/psig)
• Cylinder duration (hours) = (Cylinder volume ÷ INOMAX flow rate) ÷ 60
Calculations are considered estimates and may vary under clinical
circumstances. For more information, call 1-877-KNOW-INO (1-877-566-9466).
5 L/min
10 L/min
39 Days
19.5 Days
19.4 Days
9.7 Days
9.6 Days
4.8 Days
4.7 Days
2.3 Days
2.2 Days
1.1 Days
and doses higher than 20 ppm are not
IR
FLOW
20 L/min
9.8 Days
4.8 Days
2.4 Days
1.2 Days
13.3 Hours 6.6 Hours
15
Duration Chart
(88-size)
40 L/min
4.9 Days
2.4 Days
1.2 Days
14 Hours
Part No. 20751 Rev-01