Duration Chart Inomax Cylinder 88-Size - Mallinckrodt INOmax DSIR Plus MRI Operation Manual

Table of Contents

Advertisement

Duration Chart
INOMAX Cylinder 88-Size
For an 88-Size 800 ppm Cylinder Concentration*
(Illustrative Only)
5 L/min
5
39 Days
10
19.4 Days
20
9.6 Days
40
4.7 Days
80
2.2 Days
This chart is representative of a range of doses available on the INOmax DS
than 20 ppm are not the recommended therapeutic dose.
* All calculations in the table above are based on a full cylinder, 128 bar (1850 psig), 1,963 liter "88" cylinder,
with a cylinder change at 14 bar (200 psig). The figures are calculated based on a total continuous breathing
circuit gas flow and a cylinder conversion factor of 15.7 liters per bar/1.08 liters per psig.
• INOMAX flow = [Desired dose × total ventilator flow] ÷ [Cylinder concentration - desired dose]
• Cylinder volume = Cylinder conversion factor × cylinder pressure (bar/psig)
• Cylinder duration (hours) = (Cylinder volume ÷ INOMAX flow rate) ÷ 60
Calculations are considered estimates and may vary under clinical circumstances.
For more information, call 1-877-KNOW-INO (1-877-566-9466)
3-14
FLOW
10 L/min
20 L/min
19.5 Days
9.8 Days
9.7 Days
4.8 Days
4.8 Days
2.4 Days
2.3 Days
1.2 Days
1.1 Days
13.3 Hours 6.6 Hours
40 L/min
4.9 Days
2.4 Days
1.2 Days
14 Hours
Plus MRI and doses higher
IR
Part No. 20568 Rev-02
2015-08