Functional Safety; Appropriate Monitoring - Dräger Oxylog 3000 plus Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

For Your Safety and that of Your Patients

Functional safety

The essential performance of the Oxylog 3000 plus
is defined as:
Appropriate delivery of ventilation to the patient-
connection port or generation of an alarm condition.

Appropriate monitoring

CAUTION
Always use a separate SpO
patients who are dependent on an exact O
concentration.
The monitoring functionality of the
Oxylog 3000 plus ensures appropriate monitoring
of ventilation therapy. To ensure appropriate
monitoring during ventilation, always set the
following alarm limits:
– Airway pressure, Paw
– Expiratory minute volume, MVe
– Respiratory rate (if applicable), RR
– etCO
(if applicable)
2
Not setting appropriate alarm limits could suppress
alarms related to:
– Acute changes in the patient's condition
– Incorrect settings and faulty handling
– Hose system leakage
10
monitor for
2
2
Instructions for Use Oxylog 3000 plus SW 1.n