Operation
NIV – Non-invasive ventilation (Mask ventilation)
Use of NIV
NIV can only be activated as a supplementary
function in the ventilation modes SpnCPAP (/PS),
PC-BIPAP (/PS), VC-CMV / AF, VC-AC / AF and
VC-SIMV / AF. The Oxylog 3000 plus automatically
adjusts to the requirements of mask ventilation.
Mask leakages are detected by the device and
compensated for. Therefore, the displayed
measured values VTe and MVe do not include the
leakage. The leakage alarm is inactive.
WARNING
If NIV is not activated, measured values for VTe
and MVe will be inconsistent if there are
leakages during ventilation.
WARNING
Ensure that NIV is not activated for intubated
patients.
Risk of undetected leakages and inadequate
ventilation.
WARNING
Bacterial filters, HME and masks increase the
resistance and dead space volume of the
ventilation equipment. Note the
manufacturer's directions.
Risk of CO
rebreathing.
2
WARNING
Check MVe alarm limits after deactivating NIV
mode.
WARNING
Avoid high airway pressure.
Risk of aspiration.
76
To switch on NIV
1 Activate NIV off.
2 Select NIV on and confirm.
– The supplement NIV appears in the ventilation
mode window.
WARNING
Set the lower alarm limit MVe
the minimum ventilation required for the
patient.
Otherwise, there is a risk of the patient
receiving insufficient ventilation.
NOTE
Apnoea ventilation is not possible when NIV is
active.
Instructions for Use Oxylog 3000 plus SW 1.n