Supply From A Piped O2 System - Dräger Oxylog 3000 plus Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

Assembly
2 Connect the O
hose (A) to the
2
Oxylog 3000 plus.
3 Connect the O
hose to the pressure reducer
2
(B).
4 Rotate the cylinder valve (C) slowly and open
fully.
WARNING
Do not connect flow control valves or flowme-
ters in the gas supply to the Oxylog 3000 plus.
The ventilator could malfunction.
WARNING
Always check the O
2
before use, to prevent insufficient oxygen
supply during use.
42
A
C
B
O
2
pressure of cylinder
Supply from a piped O
A
1 Connect the O
hose (A) to the
2
Oxylog 3000 plus.
2 Connect the gas probe (B) to the O
unit until it has properly engaged and the supply
of O
is assured.
2
Instructions for Use Oxylog 3000 plus SW 1.n
system
2
B
terminal
2