Dräger Oxylog 3000 plus Instructions For Use Manual page 55

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

Switch the device ON
1 To switch the device ON briefly press
the
key (A).
The device performs a selftest and the operator is
prompted, on the display, to activate the
configuration menu or device check:
Press rotary knob for device check and
configuration
Instructions for Use Oxylog 3000 plus SW 1.n
A
Oxylog 3000 plus
B
2 Press the rotary knob (B) to confirm, before the
black progess bar is complete. The start-up
screen appears:
3 Select Device check in the start-up menu and
confirm.
NOTE
The device check can be discontinued at any time
by pressing the Alarm Reset key.
NOTE
During device check, the connections (gas supply,
hose type) and the system (flow, pressure levels,
alarm signals and knobs) are checked.
Getting Started