Spncpap, Spncpap/Ps - Dräger Oxylog 3000 plus Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

SpnCPAP, SpnCPAP/PS

Spontaneous Continuous Positive Airway
Pressure
WARNING
Only use SpnCPAP (/PS) for patients with
sufficient spontaneous breathing.
Otherwise there is a risk of the patient
receiving insufficient ventilation.
Set the ventilation pattern with the controls below
the display:
– Maximum airway pressure Pmax.
– O
concentration FiO
2
The following can be set on the display:
– Positive end expiratory pressure PEEP.
Δ
– Pressure Support
– NIV - Non-invasive ventilation.
Refer to "NIV – Non-invasive ventilation (Mask
ventilation)" on page 76.
Instructions for Use Oxylog 3000 plus SW 1.n
.
2
Psupp above PEEP.
– Insp.term.%PIF
Inspiration termination criteria of pressure sup-
ported strokes, as percentage of the peak
inspiratory flow (PIF).
– Hose type
The selected hose type must match the hose
type in use. Otherwise a correct volume mea-
surement cannot be guaranteed.
– Cuvette type (optional)
NOTE
The cuvette windows of the reusable cuvette and
disposable cuvette have different optical
properties. Therefore, the correct cuvette type
must be selected in the Setting menu. Otherwise
the zero point is shifted by up to ± 8 mmHg of CO
Operation
.
2
73