Operating Data; Power Supply - Dräger Oxylog 3000 plus Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

Technical Data

Operating data

Power supply

Power supply
Current consumption
Battery type
Charging time
Permissible ambient temperature during charging
Indication of battery capacity
Accuracy of the capacity indication
146
Oxylog 3000 plus
Input voltage
Power supplies (AC/DC
power pack and DC/DC
converter) are specified
as part of the
Oxylog 3000 plus.
With battery charging
Typical
Operating time with fully
charged internal battery
without mains power sup-
ply for "typical" ventilation
24 V ±6 VDC
maximum 2.4 A at 19 VDC
2.1 A
approximately 4 hours
Lithium ion battery
Approx. 5 hours
The specified charging time applies when recharging
the battery completely after it has been depleted.
Charging of a completely depleted battery is only
possible when the ventilator is switched OFF.
0 °C to 35 °C
in 25 % increments.
The indicated capacity is determined by the battery
itself. The accuracy depends on the type and
manufacturer and may deteriorate with frequent
partial discharge and during operation in extreme
temperatures. The internal battery is only
reconditioned after being discharged completely and
recharged at room temperature 25 °C.
The criteria for the warnings !!! Int. battery
discharged and !! Charge int. battery are therefore
based on measurement of the battery voltage. The
capacity indicated at this moment may differ from the
actual capacity of the internal battery.
Instructions for Use Oxylog 3000 plus SW 1.n