SpO2 Measurement (Optional)

SpO
Measurement (Optional)
2
Selecting a Sensor
Only Nellcor sensors may be used (see the
Accessories List, P/N 8603528) for list of available
sensors).
The new Oximax modules implemented in the Apollo
are only compatible with the new Oximax sensors
(purple probe or white probe for MAX FAST).
Only the DEC-8 or DEC-4 extension lead (purple
plug connector) may be used.
The new sensors are backward-compatible with all
modules already used in the field in older Dräger
machines.
WARNING !
Follow the Instructions for Use for the SpO
sensor being used. Incorrect positioning may
cause injuries.
The following table provides guidelines for selecting
a specific sensor.
OXIMAX™
Sensor type:
MAX N
Neonates/
Age group
Adults
<3 to >40 kg
Weight
of the patient
(<6.6 to >88 lb)
Short and long-term monitoring
Duration
of use
Mobility
of the patient
Ball of the foot Toe
Preferred
measuring
point
Sterile
a
Sterility
packaging
a. In unopened and undamaged packaging
DURASENSOR™ and OXIMAX™ are registered trademarks.
Apollo Operator's Manual
2
OXIMAX™
OXIMAX™
MAX I
MAX P
Infants
Children
1 to 20 kg
10 to 50 kg
(2.2 to 44 lb)
(22 to 110 lb)
Limited activity
Sterile
Sterile
packaging
packaging
DURASENSOR™
OXIMAX™
DS-100 A
MAX A
>40 kg
>30 kg
(>88 lb)
(>66 lb)
Short-term
monitoring
Inactive patients
Limited activity Inactive
only
Finger
–––––––
Sterile
packaging
Chapter 8 - Monitoring
OXIMAX™
OXIMAX™
MAX R
MAX FAST
Adults
>50 kg
>40 kg
(>110 lb)
(>88 lb)
Short and long-term monitoring
Limited
patients only,
activity
must be
checked at
least every 8
hours
Nose
Forehead
Sterile
Sterile
packaging
packaging