Reprocessing list
Applicable to non-infectious patients.
The list contains approximate values only. The
Instructions for Use of the hospital's infection con-
trol officer shall prevail and must be observed by
the user!
Components which
can be reprocessed
Oxylog 3000 plus
device and O
hoses
2
Reusable breathing
circuit
CO
sensor
2
Reusable cuvette of
the CO
sensor
2
Test filter for CO
sen-
2
sor
Assembling parts
– Reassemble, refer to "Assembly" on page 35
for information.
– Connect to the power supply and gas supply,
refer to "Assembly" on page 35 for information.
– Check readiness for operation, refer to "Getting
Started" on page 51 for information.
Instructions for Use Oxylog 3000 plus SW 1.n
Recommended
Machine
reprocessing
cleaning and
intervals
disinfection
Per patient /
No
if soiled
Per patient /
Yes
if soiled
Per patient /
No
if soiled
Per patient /
Yes
if soiled
If soiled
No
Cleaning, Disinfection and Sterilization
Manual
Cleaning
Disinfection
Outside
Outside only
only
Possible
Possible
Outside
Outside only
only
Yes
Yes
Yes
Yes
Sterilization
No
Yes
No
Yes
No
131