Performance Data - Dräger Oxylog 3000 plus Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

Performance data

Control principle
Maximum inspiratory flow
Device compliance
Device inspiratory and expiratory resistance with
adult hose systems
Device inspiratory and expiratory resistance with
paediatric disposable hose systems
Dead space including flow sensor, but excluding
accessories such as filters, HMEs
and CO
cuvettes
2
Dead space CO
cuvettes
2
Resistance CO
cuvettes
2
Supplementary functions
Patient connection
1) At service pressures >350 kPa.
The maximum inspiratory flow is reduced to 80 L/min at service pressures <350 kPa and to 39 L/min at service
pressures <280 kPa.
2) When using an accessory with a female connector, add 2 mL to the hose system dead space.
Instructions for Use Oxylog 3000 plus SW 1.n
with 1.5 m adult
breathing hose
with 3 m adult
breathing hose
with 1.9 m paediatric
breathing hose
Demand valve
Relief valve
Time-cycled, volume-constant, pressure-controlled
1)
100 L/min
<1 mL/mbar
<2 mL/mbar
<0.7 mL/mbar
<6 mbar at 60 L/min
<4 mbar at 30 L/min
<2 mbar at 5 L/min
<5 mbar at 30 L/min
<2 mbar at 5 L/min
2)
approximately 35 mL
system).
2)
approximately 30 mL
system).
2)
approximately 15 mL
system).
approximately 4 ml (adult CO
approximately 1.5 mL (paediatric CO
<1 mbar at 60 L/min. (adult CO
<5 mbar at 30 L/min. (paediatric CO
Opens the breathing system upon failure of the gas
supply, permits spontaneous breathing with ambient
air.
Opens the breathing system in case of device failure
at approximately 80 mbar.
22 mm ISO conical connector
Technical Data
(reusable adult hose
(disposable adult hose
(disposable paediatric hose
cuvette).
2
cuvette).
2
cuvette).
2
cuvette).
2
143