Manual disinfection
Manual disinfection should preferably be carried
out with disinfectants based on aldehydes or quar-
ternary ammonia compounds.
Observe the applicable country-specific listings for
disinfectants. The list of the German Association for
Applied Hygiene (Verbund für Angewandte
Hygiene VAH) applies in German-speaking coun-
tries.
The composition of the disinfectant is the responsi-
bility of the manufacturer and can change over
time.
Strictly observe the manufacturer's information on
the cleaning agent.
Carry out manual disinfection
1 Disinfect items by immersing bubble-free for
approx. 30 minutes.
2 Carry out manual cleaning as described in step
2 of "Carry out manual cleaning" on page 128.
3 After contact time, rinse items under running
water until disinfectant residue is no longer dis-
cernible.
4 Inspect items for visible soiling and damage. If
necessary, repeat manual disinfection.
5 Thoroughly shake out residual water and allow
items to dry completely.
Machine cleaning and disinfection
Use washer-disinfector in accordance with
EN ISO 15883, preferably with a cart for anesthe-
sia and ventilation accessories, for automatic
cleaning and disinfection. Use mildly alkaline or
enzymatic (with neutral pH) cleaning agent. Strictly
observe the manufacturer's information on the
cleaning agent.
Instructions for Use Oxylog 3000 plus SW 1.n
Cleaning, Disinfection and Sterilization
Carry out machine cleaning and disinfection
1 Strictly observe Instructions for Use of washer-
disinfector.
2 Position items so that all interior spaces are
completely flushed and water can drain off
freely.
– Connect hoses to suitable nozzle of the car-
riage in the washer-disinfector and place
them evenly.
– Connect breathing valve (without cover and
silicone diaphragm) with a short breathing
hose (e.g., 35 cm silicone breathing hose,
order no. 2165619) to a suitable nozzle of
the carriage in the washer-disinfector.
3 Use suitable cleaning agent.
4 Select suitable program (preferably anesthesia
program).
– Cleaning must be carried out at 40 °C to
60 °C (104 °F to 140 °F) for at least
5 minutes.
– Thermal disinfection must be carried out at
80 °C to 95 °C (179 °F to 203 °F) and with
corresponding contact time.
5 Carry out final rinsing with deionized water.
6 Immediately remove items from the washer-dis-
infector.
7 Inspect items for visible soiling and damage. If
necessary, repeat program or carry out manual
cleaning and disinfection.
8 Allow items to dry thoroughly.