Operation
When patients are breathing spontaneously, the
achievable O
concentration will depend on the
2
profile of the inspiratory flow.
WARNING
In toxic surroundings:
– The patient must be ventilated with 100 %
medical grade oxygen so that toxic constit-
uents do not enter into the breathing gas.
– The patient must be immediately trans-
ferred to a breathable atmosphere in order
to prevent inhalation of toxic air when
spontaneous breathing resumes.

Setting HME correction

The temperature and moisture influence of HME
have an effect on the flow measurement. The
Oxylog 3000 plus can compensate for the
presence of an HME.
When using HME, select, set and confirm HME -
On in the Settings window with the rotary knob.
When HME - ON is selected, the flow sensor
expects an expiration gas temperature of 35 °C and
a relative humidity of 0 %.
When HME - Off is selected, the flow sensor
expects an expiration gas temperature of 37 °C and
a relative humidity of 100 %.
80
WARNING
In infectious environments:
– The patient must be ventilated with 100 %
medical grade oxygen so that bacteria,
viruses, fungi or spores do not enter the
breathing gas.
– The patient must be immediately trans-
ferred to a breathable atmosphere in order
to prevent inhalation of infectious air when
spontaneous breathing resumes.
Instructions for Use Oxylog 3000 plus SW 1.n