Ambient Conditions; During Operation; During Storage And Transportation - Dräger Oxylog 3000 plus Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

Technical Data
WARNING
Do not use the device outside the specified
environmental and supply conditions, as the
device may not operate according to its
specifications and may become inoperative.

Ambient conditions

During operation

During storage and transportation

Ventilator without replaceable battery,
with reusable hose system
Disposable adult and paediatric hose sys-
tems
Replaceable battery
140
Temperature
Temperature range for
CO
sensor
2
Atmospheric pressure 570 to 1200 hPa
Automatic atmospheric pressure compensation within operating pressure range
Operating pressure for
AC/DC power pack
Operating pressure for
CO
sensor
2
Relative humidity
Temperature
Atmospheric pressure 570 to 1200 hPa
Atmospheric pressure
for CO
sensor
2
Relative humidity
Temperature
Atmospheric pressure 570 to 1200 hPa
Relative humidity
Temperature
Atmospheric pressure 570 to 1200 hPa
Relative humidity
–20 to +50 °C
+10 to +40 °C
Usage outside this temperature range may decrease
sensor accuracy. Usage below +10°C increases the heat
up time.
700 to 1200 hPa
570 to 1100 hPa
5 to 95 % (no condensation)
–40 to +75 °C
115 to 1100 hPa
5 to 95 % (no condensation)
–20 to +70 °C
5 to 95 % (no condensation)
–20 to +35 °C
5 to 95 % (no condensation)
Instructions for Use Oxylog 3000 plus SW 1.n