Operation

Shutdown

– After disconnecting the patient
Switch the device OFF:
1 To switch the device OFF, hold down the
key (A) for approximately 3 seconds.
Ventilation is now stopped and a high-priority
alarm is issued.
This alarm can be silenced with the
2 Either:
– Press the rotary knob (B) to confirm switch
off.
Or
– Press the
key (A) to resume ventilation
with the previous settings.
NOTE
When the device is switched OFF, the battery is
still being charged if the device is connected to an
external power source.
82
When O
3 Close the cylinder valve.
WARNING
The cylinder valve must be closed completely
to avoid gas flow leakage by the device.
When O
Oxylog 3000 plus
4 Disconnect the high-pressure connection from
5 Disconnect the breathing hose.
B
A
key.
CAUTION
When disconnecting the breathing hose, always
grip the sleeve (A) and not the corrugations (B).
If this is not done, the corrugations or hose may
be torn from the sleeve.
is supplied from a cylinder:
2
is supplied from the pipeline system:
2
the source.
A
B
Instructions for Use Oxylog 3000 plus SW 1.n