Internal rechargeable battery

Internal power is provided by means of a
removable rechargeable battery. For technical
information, refer to "Technical Data" on page 139.

Removing the battery

A
B
C
1 Turn the knob (C) on the battery compartment
cover (B) counterclockwise to release the cover.
2 Open the battery cover.
3 Remove the battery (A) by pulling the tab.
Checking the charge status of the
battery
Press the button on the rechargeable battery.
The charge status is indicated as a percentage
by an indicator.
Instructions for Use Oxylog 3000 plus SW 1.n

Installing the battery

1 Insert the battery into the battery compartment.
2 Close the battery cover.
3 Tighten the knob by turning it.
CAUTION
The Oxylog 3000 plus will interrupt ventilation
when the battery is replaced while the device is
switched on and the external power supply is not
connected. Ventilation will always resume with the
last values settings approximately 3 seconds after
inserting a recharged battery.
Assembly