SpO
measurement (optional)
2

Safety and precautions

– The sensor must not be applied to limbs together with
an arterial catheter, infusion or sphygmomanometer
cuff.
– Blood circulation must not be impeded during
application of the sensor. If possible, the site should
be changed from time to time in order to avoid
pressure necroses at the measuring point.
– The sensor should be protected against bright light
(covered).
– Nellcor sensors should be used exclusively and
applied as described below.
– Damaged sensors must not be used.
– The adhesive strip of the Oxiband sensor must be
discarded after use. It should not be stretched unduly.
Never use two strips together.

Choice of sensor

Nellcor sensors must be used exclusively. Note the
Instructions for Use of the sensors. Tissue damage may
result if they are positioned or used incorrectly.
The choice of sensor should be based on the following
criteria:
– Patient's weight
– Mobility of the patient
– Possible application site
– Perfusion of the patient
– Period of use
The following table listing the various sensors available
and their specific characteristics may prove useful here.
Type of
OXISENSOR
sensor
I-20
Age group
Infants
Weight of
1 to 20 kg
patient
Period of use Short and long-term
monitoring
Mobility of
Limited activity
patient
Preferred site Toe
1)
Sterility
In sterile packaging In sterile packaging
,
OXISENSOR
I-20
OXISENSOR
DURASENSOR
DS-100A
OXISENSOR
R-15 are registered trademarks.
1) in undamaged, unopened packaging.
OXISENSOR
D-20
Children
10 to 50 kg
Short and long-term Short-term
monitoring
Limited activity
Finger
,
D-20
,
OXISENSOR
D-25 and
DURASENSOR OXISENSOR
DS-100A
D-25
Adults
Adults
> 40 kg
> 30 kg
Short and long-term Short and long-term
monitoring
monitoring
Inactive patients Limited activity
only
Finger
Finger
____
In sterile packaging In sterile packaging
Care
Reassembling system
SpO
measurement
2
Safety and precautions
Choice of sensor
OXISENSOR
R-15
Adults
> 50 kg
monitoring
Inactive patients
only
Nose
95