Dräger Oxylog 3000 plus Instructions For Use Manual page 154

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

Technical Data
Immunity against
Power frequency
magnetic field
50/60 Hz
(IEC 61000-4-8)
Voltage dips and
short interruptions
on AC mains input
lines
(IEC 61000-4-11)
Radiated RF
(IEC 61000-4-3)
RF coupled into lines
(IEC 61000-4-6)
Information regarding separation distances (IEC 60601-1-2, tables 5 and 6).
X1)
For P
the highest possible "equivalent isotropic radiated power" of the adjacent RF transmitter has to be inserted
EIRP
(value in Watt). Also in the vicinity of equipment marked with the symbol
fixed, portable or mobile RF transmitters at the location of the Oxylog 3000 plus should be less than 3 V/m in the frequency
range from 150 kHz to 2.5 GHz and less than 1 V/m above 2.5 GHz.
X2
)
ISM bands in this frequency range are: 6.765 MHz to 6.795 MHz, 13.553 MHz to 13.567 MHz,
26.957 MHz to 27.283 MHz, 40.66 MHz to 40.70 MHz.
154
IEC 60601-1-2 test level
3 A/m
dip >95 %, 0.5 periods
dip 60 %, 5 periods
dip 30 %, 25 periods
dip >95 %, 5 seconds
80 MHz to 2.5 GHz: 10 V/m 30 V/m
150 kHz to 80 MHz: 10 V
within ISM bands,
3 V
outside ISM bands
rms
(X2)
Compliance
level (of the
Oxylog 3000
plus)
3 A/m
>95 %,
0.5 periods
60 %, 5 periods
30 %,
25 periods
>95 %,
5 seconds
10 V
rms
rms
10 V
rms
interference may occur. Field strengths from
Instructions for Use Oxylog 3000 plus SW 1.n
Electromagnetic environment
Test is not necessary because
the Oxylog 3000 plus is not likely
to be sensitive to magnetic field
disturbances, such as CRT mon-
itors or hall elements.
No voltage dips or short interrup-
tions, since a fully charged bat-
tery must be installed for safety
reasons, even when operating
from an external supply.
Recommended separation dis-
tance from portable and mobile
RF transmitters with transmis-
sion power P
to the
EIRP
Oxylog 3000 plus including its
lines:
(X1)
1.84 m *
P
EIRP
NOTE
The CO
sensor has a lower
2
compliance level (20 V/m) but
will fail in a safe way.
Recommended separation dis-
tance from portable and mobile
RF transmitters with transmis-
sion power P
to the
EIRP
Oxylog 3000 plus including its
lines:
1.84 m *
P
EIRP
(X1)