Test Loudspeaker, Buzzer, Leds And Display; Display Battery And Supply Data - Dräger Oxylog 3000 plus Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

Set the controls accordingly for the test:
– VT to 500 mL
– RR to 20/min
– Pmax to 40 mbar
– FiO
to 40 %
2
These settings are displayed on the screen.
To test the buttons:
1 Briefly press the corresponding button.
The associated letter on the screen changes
from "B" to "X". If the button has an indicator, it
will be illuminated by the device. If there are but-
tons without an indicator, the yellow warning
indicator will light up on the device.
NOTE
If you press the
key for longer than 3 seconds,
the ventilator switches OFF.
The function of the rotary knob is not included in the
test.
Test loudspeaker, buzzer, LEDs and
display
To test the loudspeaker, buzzer, all LEDs and the
display:
1 Select the required test.
2 Confirm the test with the rotary knob. The
requested function is tested by the device.
To test the screen display (Test display).
Instructions for Use Oxylog 3000 plus SW 1.n
3 Turn the rotary knob; various test cards are
displayed.
The selected test remains active until the rotary
knob is pressed again.

Display battery and supply data

The parameters of the replaceable battery and the
status of the external power supply are displayed.
Display (example):
Advance to the second page:
Select the line Page, confirm and turn rotary
knob.
Display (example):