Problem Solving

Messages in the information window

The numeric values below are examples.
Message
Confirm PEEP above 10 mbar ?
Gas consumption = 10 L/min
(Battery capacity)
REUS = Adult reusable hose
DISP = Adult disposable hose
PAED = Paediatric disposable
hose
Select and confirm connected
hose with rotary knob.
REUS = Reusable Cuvette
DISP = Disposable Cuvette
Press Rotary Knob for Autoset
/ Exit with Alarm Reset
Remove sensor from cuvette
then press Rotary Knob
Place sensor on reference filter
then press Rotary Knob
Pinsp>=PEEP + 3 mbar
120
1)
Unit
Cause
PEEP >10 mbar has been set but
not confirmed.
Standard display in information
window for the current gas
consumption.
Standard display in information
window for the current battery
capacity.
Explanation of abbreviation,
when selecting the hose type in
the Settings window.
Confirm hose type.
Explanation of abbreviation,
when selecting the cuvette type in
the Settings window.
Activation of Alarms: Autoset,
refer to "Setting alarm limits"
on page 87.
Activation of CO
tion, refer to "CO
(optional)" on page 91.
Activation of CO
refer to "CO
measurement
2
(optional)" on page 91.
Set PEEP + 3 mbar > Pinsp
Explanation/Remedy
The required setting of
PEEP >10 mbar is only
possible when confirmed
via the rotary knob.
Select and confirm.
After pressing the rotary
knob the new alarm limits
will be set.
zero calibra-
After removal of the CO
2
measurement
sensor from the cuvette and
2
pressing the rotary knob,
the zero calibration will pro-
ceed. Cancel by pressing
Alarm Reset.
filter check,
After mounting the test filter
2
and pressing the rotary
knob, the filter check will
proceed. Cancel by press-
ing Alarm Reset.
Set Pinsp > PEEP + 3 mbar
Instructions for Use Oxylog 3000 plus SW 1.n