Operation
– Hose type
The selected hose type must match the hose
type in use. Otherwise a correct volume
measurement cannot be guaranteed.
– Cuvette type (optional)
NOTE
The cuvette windows of the reusable cuvette and
disposable cuvette have different optical proper-
ties. Therefore, the correct cuvette type must be
selected in the Setting menu. Otherwise the zero
point is shifted by up to ± 8 mmHg of CO
66

Trigger (VC-AC)

NOTE
When in VC-CMV and the trigger is set to a value,
the ventilation mode changes to VC-AC.
VC-AC –Volume Controlled - Assist Control
For synchronization with the patient's spontaneous
inspiratory efforts.
The mandatory breaths are synchronized with the
patient's spontaneous inspiratory efforts when the
trigger is activated and the trigger sensitivity is set.
In this case, the actual respiratory rate may be
higher than the set ventilation respiratory rate RR.
Successful patient triggering is briefly indicated by
an asterisk (*) on the left side of the curve window.
.
2
Instructions for Use Oxylog 3000 plus SW 1.n