Connecting The Paediatric Disposable; Hose System - Dräger Oxylog 3000 plus Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

Assembly
Connecting the paediatric disposable hose system
B
A
1 Connect the blue flow measuring line (B) to the
blue labeled connector.
2 Connect the transparent flow measuring line (A)
to the other connector.
3 Connect the breathing hose (C) to the gas outlet
on the Oxylog 3000 plus.
When connecting a hose, check that the hose
setting in the Settings window corresponds to the
connected hose.
46
C
Instructions for Use Oxylog 3000 plus SW 1.n