Operation

VC-SIMV, VC-SIMV/PS

Volume Controlled - Synchronized Intermittent
Mandatory Ventilation
Paw
P
PEEP
T
i
1
Flow
RR
Insp. Flow
For patients with inadequate spontaneous breath-
ing, or for patients who are to be weaned gradually.
Fixed mandatory minute volume MV is set with tidal
volume VT and ventilation respiratory rate RR. The
patient can breathe spontaneously between the
mandatory breaths and thus contribute to the total
minute volume. Spontaneous breathing can be
assisted with PS.
Set the ventilation pattern with the controls below
the display:
– Tidal volume VT.
– Respiratory rate RR.
(minimum possible respiratory rate: 2 per min).
– Maximum airway pressure Pmax.
– O
concentration FiO
2
68
Pressure
support PS
max
short
long
rise time
rise time
Trigger window
.
2
The following can be set on the display:
t
– Positive end expiratory pressure PEEP.
– Pressure Support
– Sensitivity Trigger.
t
Successful patient triggering is indicated by an
asterisk (*) on the left side of the curve window.
– Ventilation time ratio I:E or inspiratory time Ti.
When setting the ventilation respiratory rate RR,
tidal volume VT or I:E / Ti, the associated values for
inspiration flow and Ti / I:E are automatically dis-
played in the information window.
– Plateau time Tplat %, in % of the inspiratory
time.
– AutoFlow (optional).
Instructions for Use Oxylog 3000 plus SW 1.n
Δ
Psupp above PEEP.