Set Start-Up Settings - Dräger Oxylog 3000 plus Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

Configuration

Set start-up settings

The ranges of the settings are:
Parameter
Range
Ventilation mode
Trigger
0 (OFF) 1 to 15 L/min
PEEP
0 to 20 mbar
I:E or Ti
Configurable I:E or Ti
I:E
1:100 to 50:1
Ti
0.2 to 10.0 s
Tplat %
0 to 50 %
ΔPsupp
0 to 35 mbar
Slope
FLAT, MEDIUM, STEEP
Pinsp
3 to 55 mbar
O
-Flow
0 to 15 L/min
2
NIV
ON, OFF
Tapn
0 (OFF) to 60 s
VTapn
50 to 2000 mL
RRapn
12 to 60 bpm
MVe-high alarm
2.0 to 41 L/min
MVe-low alarm
0.5 to 40 L/min
RR-high
10 to 100 bpm
HME correction
ON, OFF
AutoFlow
ON, OFF
Brightness-min*
1/4 to 4/4
Brightness
1/4 to 4/4
Alarm volume
1/4 to 4/4
etCO
high alarm
1 to 100 mmHg
2
etCO
low alarm
0 to 100 mmHg
2
CO
unit
mmHg, kPa or Vol.%
2
Hose type
Adult disposable
Adult reusable
Paediatric disposable
Cuvette type
Disposable
Reusable
* Brightness-min: screen brightness level in power
save mode. Refer to "Screen brightness"
on page 81.
100
The manufacturer's default settings are:
Parameter
Manufacturer's default setting
Ventilation mode VC-CMV
Trigger
0 L/min at VC-CMV, VC-AC
and 3 L/min at VC-SIMV,
SpnCPAP, PC-BIPAP
PEEP
5 mbar
I:E or Ti
I:E
I:E
1.0:1.5
Ti
2.0 s
Tplat %
0 %
ΔPsupp
0 mbar
Slope (pressure
Standard
rise time)
Pinsp
20 mbar
O
-Flow
10 L/min
2
NIV
OFF
Tapn
0 s (when switched on
minimum of 15 s)
VTapn
500 mL
RRapn
12 /min
MVe-high alarm
40.0 L/min
MVe-low alarm
0.5 L/min
RRsp-high alarm 100 /min
HME correction
OFF
AutoFlow
OFF
Brightness-min*
1/4
Brightness
3/4
Alarm volume
3/4
etCO
high alarm 50 mmHg
2
etCO
low alarm 33 mmHg
2
CO
unit
mmHg
2
Hose type
Adult disposable
Cuvette type
Disposable
Instructions for Use Oxylog 3000 plus SW 1.n