Settings - Dräger Oxylog 3000 plus Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

Settings

Ventilation modes
Respiratory rate RR
Ventilation time ratio I:E
Inspiratory time Ti
Tidal volume VT
Inspiratory pressure Pinsp
O
concentration
2
Positive end expiratory pressure PEEP
Trigger sensitivity (flow trigger)
Pressure Support ΔPsupp
Alarm limit ranges
Instructions for Use Oxylog 3000 plus SW 1.n
Accuracy of Setting
Accuracy of Setting
Rise time for Pressure
Support
Alarm
MVe
MVe
VC-CMV, VC-AC, VC-SIMV, VC-SIMV/PS, SpnCPAP,
SpnCPAP/PS, PC-BIPAP, PC-BIPAP/PS
Optional: AutoFlow for VC-CMV, VC-AC and
VC-SIMV (/PS)
2 to 60 /min (VC-SIMV, PC-BIPAP)
5 to 60 /min (VC-CMV, VC-AC)
12 to 60 /min for apnoea ventilation
1:100 to 50:1
0.2 to 10 s
0.05 to 2.0 L, BTPS
Measured values referred to the conditions of the
patient's lungs, body temperature 37 °C, airway pres-
sure, water-vapor-saturated gas.
±15 % of set value or ±25 mL, whichever is greater
(adult breathing hose).
±15 % of set value or ±15 mL, whichever is greater
(paediatric breathing hose).
PEEP +3 to +55 mbar
40 to 100 Vol.%
±10 Vol.% The actual value depends on the inspiratory
1)
flow
and mean airway pressure.
0 to 20 mbar, no negative pressure.
1 to 15 L/min
0 to 35 mbar (relative to PEEP)
long (1 s), standard (0.4 s), short.
Alarm limit range
2 to 41 L/min
0.5 to 40 L/min
Technical Data
141