WARNING
A potential hazard may exist if different
default alarm settings are configured for the
same or similar equipment in any single area,
e.g. an emergency department.
The user default settings for the parameters are
displayed on the screen when the ventilator is
switched ON. The settings can be adjusted.
Advance to the second page:
1 Select the line Page, confirm and turn rotary
knob.
To restore the manufacturer's defaults:
2 Select and confirm line Set factory default.
Instructions for Use Oxylog 3000 plus SW 1.n

Hose/cuvette start-up settings

– Ask for hose type = Yes: When switching on the
device, prior to ventilation, the user is first asked
to select the hose type.
– Ask for hose type = No: When switching on the
device, it directly starts ventilating after the self-
test, with the default hose type setting.
– Hose type: The start-up setting for the breathing
hose (Adult Reusable, Adult Disposable or Pae-
diatric Disposable).
– Cuvette type (optional): The start-up setting for
the CO
cuvette type (Reusable or Disposable).
2
Configuration
101