VC-CMV, VC-AC

VC-CMV -Volume Controlled - Controlled
Mandatory Ventilation
Paw
Plateau time
Tplat
P
plat
T
T
insp
e
1
RR
Flow
Insp. Flow
Exp. Flow
Volume-controlled ventilation with fixed mandatory
minute volume MV, which is set with tidal volume
VT and respiratory rate RR.
WARNING
Only use VC-CMV for patients who are not
spontaneously breathing. Otherwise, the
patient may be put at risk by not receiving
sufficient ventilation.
Use VC-AC for patients with partial
spontaneous breathing.
Set the ventilation pattern with the controls below
the display:
– Tidal volume VT.
– Ventilation respiratory rate RR.
(minimum possible frequency: 5 per min).
– Maximum airway pressure Pmax.
– O
concentration FiO
2
Instructions for Use Oxylog 3000 plus SW 1.n
P
max
PEEP
.
2
The following can be set on the display:
t
– Positive end expiratory pressure PEEP.
– Sensitivity Trigger.
– Ventilation time ratio I:E or inspiratory time Ti.
t
When setting the ventilation respiratory rate RR,
tidal volume VT or I:E / Ti, the associated values for
inspiration flow and Ti / I:E are automatically
displayed in the information window.
– Plateau time Tplat %, in % of the inspiratory
time.
– AutoFlow (optional).
Operation
65