Dräger Oxylog 3000 plus Instructions For Use Manual page 56

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

Getting Started
Check connections
1 Ensure that the gas supply has been
connected.
2 Select and confirm the appropriate hose type.
3 Ensure that the test lung has been connected.
The Oxylog 3000 plus automatically checks if a
test lung has been connected. The device
check is interrupted if a test lung is not detected
within one minute. The check is continued when
the test lung is detected.
4 The Oxylog 3000 plus automatically checks if
the detected hose differs from the selected
hose type.
If the wrong hose type is selected:
– Press the Alarm Reset key to cancel the
device check.
– Restart the device check.
– Select the correct hose type.
56
System check
A
5 Set the control knobs (A) below the display to
the required values.
The Oxylog 3000 plus successively activates
the audible and optical alarm signals and
prompts the operator to acknowledge each
signal.
6 Confirm the audible and optical alarm signals.
The device check continues automatically.
During the automatic test sequence, the
Oxylog 3000 plus checks the flow, pressure lev-
els and alarm signals. Corresponding sounds
are heard.
The bar graph shows the progress made by the
check.
The result is displayed on the last page of the
device check screens. If all tests are completed
successfully, the device will go to the last page.
If a test fails, the device will go directly after the
failed test to the last page, without performing
the other tests.
After confirmation, the system returns to the
menu screen.
If the service inspection date has been passed with-
out servicing, the text Service date overdue ! will
appear in the window after finishing the device
check. In this case the device must be serviced
immediately.
Instructions for Use Oxylog 3000 plus SW 1.n
Oxylog 3000 plus