Special functions

Manual inspiration / Inspiration hold

The function Manual inspiration / Inspiration hold
will either initiate a new (manual) breath or hold the
inspiratory phase of the current breath for a maxi-
mum of 15 seconds.
The pattern of the manually started breath corre-
sponds with the set ventilation mode.
This function is not available for:
– SpnCPAP without PS,
– O
inhalation (optional).
2
To activate Manual inspiration or Inspiration
hold
Press key Inspiration hold (A) for as long as
inspiration is required.
Instructions for Use Oxylog 3000 plus SW 1.n
A
Oxylog 3000 plus
B
100 % O
(optional)
2
To apply 100 % O
for 3 minutes regardless of the
2
momentarily set value.
Briefly press key 100 % O
Its indicator lights up for 3 minutes.
The set value is resumed by the ventilator upon
expiry of these 3 minutes, or when the 100 % O
pressed again. The indicator dims.
O
inhalation (optional)
2
WARNING
The O
inhalation function is not a ventilation
2
mode.
It may only be used for patients with spontane-
ous breathing who receive a constant O
of between 0 and 15 L/min via a mask.
If a tracheal stenosis or other obstruction occurs,
the flow is interrupted by the ventilator for
5 seconds at an airway pressure of 30 mbar and
the airway pressure is reduced to 0 mbar.
The !!! Paw high alarm is active.
NOTE
The options 100 % O
2
mutually exclusive.
Operation
(B).
2
flow
2
and O
inhalation are
2
is
2
77