Cleaning, Disinfection and Sterilization
Disassembly
Remove the CO
sensor and cuvette
2
B
A
1 Unplug the CO
sensor connector at the side of
2
the Oxylog 3000 plus.
2 Remove the CO
sensor (B) from the CO
2
cuvette (C).
3 Remove the CO
cuvette (C) from the flow
2
sensor (D).
4 Remove the angled connector (A) from the
cuvette (C).
124
D
C
2
Disassemble the adult reusable hose
system
1 Disconnect the breathing hose (A) from the gas
output.
2 Disconnect the flow measuring lines (B) from
the nozzles.
Instructions for Use Oxylog 3000 plus SW 1.n
B
A