Alarms
NOTE
The volume of alarm tones can be adjusted. Refer
to section "Alarm volume" on page 81.
If alarm tones are to be heard again before the
2 minutes have expired:
2 Press the key
(A) again.
In the event of a gas failure
WARNING
In the event of a gas failure, proper ventilation
cannot be guaranteed. The Oxylog 3000 plus
issues the alarm !!! Supply pressure low.
Disconnect the patient from the device and
continue ventilation without delay using
another ventilator.
In the event of an internal power supply failure
WARNING
In the event of an internal power supply fail-
ure, automatic ventilation, volume measure-
ment and alarms do not operate.
An acoustic alarm signal is given to indicate
the internal power supply failure.
Spontaneous breathing can continue through
the emergency air intake.
Disconnect the patient from the device and
continue ventilation without delay using
another ventilator.