To enter Customer Service Mode
E
D
A
B
1 Ensure that ventilation is OFF.
2 Turn controls VT (A) and RR (B) all the way to
the right.
3 Switch on the device by briefly pressing the
Start/Standby key
press and hold the Curves
the Values
key (E) until the main
Customer Service Mode menu appears.
4 In the main menu, set the number of the
required test with the rotary knob.
5 Press the rotary knob to activate the test.
Instructions for Use Oxylog 3000 plus SW 1.n
Oxylog 3000 plus
C
(C) and simultaneously
Curves
key (D) and
Settings in Customer Service Mode
6 Select the required function with the cursor
(asterisk).
– To select a parameter: turn rotary knob.
– To activate the parameter: press rotary
knob.
– To set a value: turn rotary knob.
– To confirm the value: press rotary knob.
To exit the parameter settings menu
1 Select the line EXIT.
2 Press the rotary knob to confirm.
The set values are saved as default settings
and remain effective.
Configuration
99