Technical Data
Measured values and curve display
Airway pressure measurement
Flow measurement
Minute volume MVe
Tidal volume VTe
CO
measurement
2
Frequency measurement
Curve display
144
Range
Resolution
Accuracy
Range
Resolution
Accuracy
Range
Resolution
Accuracy
Range
Resolution
Accuracy
Total system response
time
Range
Resolution
Accuracy
Airway pressure Paw (t) –10 to 100 mbar
Flow (t)
CO
2
0 to 100 mbar
1 mbar
±2 mbar
0 to 100 L/min, BTPS
0.1 L/min
±15 % of measured value, or ±0.4 L/min, whichever
is greater.
0 to 5000 mL, BTPS
1 mL
±15 % of measured value, or ±20 mL, whichever is
greater (adult breathing hose).
±15 % of measured value, or ±15 mL, whichever is
greater (paediatric breathing hose).
0 to 120 mmHg / 0 to 15.8 Vol.% / 0 to 16.0 kPa
1 mmHg / 0.1 Vol.% / 0.1 kPa
0.43 % by volume +8 % of the CO
per ISO 21647:2004
200 ms
0 to 99/min
1/min
±1/min
–120 to 120 L/min
–5 to 120 mmHg / –1 to 16 Vol.% /–1 to 16 kPa
Instructions for Use Oxylog 3000 plus SW 1.n
concentration as