Co2 Sensor - Dräger Oxylog 3000 plus Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

CO
sensor
2
Prerequisite: The CO
2
following actions can be selected:
– Zero Calibration;
– Filter/gas check;
– Filter check;
– Gas check;
– Gas Calibration;
– Start gas calibration;
– Reset to factory default;
Performing zero calibration
NOTE
Do not breathe on the CO
calibration, otherwise the zero calibration can fail
or the zero calibration can pass with an invalid zero
value.
Instructions for Use Oxylog 3000 plus SW 1.n
option is activated. The
sensor during zero
2
Start zero calibration:
1 Connect the CO
sensor to the ventilator. Wait
2
at least 3 minutes for the CO
complete its warm-up phase.
2 Remove the CO
sensor (A) from the
2
cuvette (B).
A
3 Open page CO
sensor, select the line Zero
2
calibration and confirm. The screen displays
the text Remove sensor from cuvette.
4 Confirm with rotary knob. The Oxylog 3000 plus
performs the zero calibration and displays the
message Zero calibration in progress.
If zero calibration was successful:
After approx. 5 seconds the Oxylog 3000 plus
confirms with the message Zero calibration
successful.
If zero calibration was not successful:
The Oxylog 3000 plus displays the alarm Zero
calibration failed.
Repeat zero calibration.
If zero calibration is still not possible:
1 Check whether the sensor (A) is soiled and
clean it if necessary. If the sensor is defective,
replace the sensor.
2 Repeat zero calibration.
Configuration
sensor to
2
B
105