Operation

Apnoea ventilation

Paw
Pmax
Apnea alarm time
Tapn
Start
Flow
apnea ventilation
with apnea alarm
Apnoea back-up ventilation is only applicable when
using the SpnCPAP mode. In the event of an
apnoea, the ventilator will automatically activate
volume-controlled mandatory ventilation - (VC-
CMV).
When an apnoea occurs, the device
simultaneously issues an alarm signal and
switches to volume controlled ventilation with the
parameters respiratory rate RRapn, tidal volume
VTapn, and the maximum airway pressure Pmax
when the apnoea time Tapn has been reached.
The ventilation time ratio I:E = 1:1.5 and the plateau
time Tplat % = 0 are preset during apnoea
ventilation.
74
Setting apnoea ventilation
t
1
RRapn
t
On the display:
The parameters RRapn and VTapn, which are
required for setting apnoea ventilation, are now
displayed:
3 Set RRapn and VTapn.
4 Set Pmax. This determines the maximum
Set Tapn with the rotary knob (B) to a value
between 15 and 60 seconds.
airway pressure allowed during apnoea
ventilation.
Instructions for Use Oxylog 3000 plus SW 1.n