Maintenance

Maintenance intervals of Oxylog 3000 plus

Definition of Maintenance Concepts
Concept
Maintenance
Inspection
Preventive maintenance
Repair
Keep a record on all preventive maintenance.
WARNING
In order to avoid malfunctioning of the device,
maintenance must be carried out by properly
trained service personnel.
WARNING
Clean and disinfect the device or device parts
before each maintenance step – and also
when returning for repair to minimize risk of
infection.
Task
1)
Replace dust filter
Replace internal battery
(refer to page 37).
Device inspection and maintenance
(including safety inspections)
CO
sensor inspection and
2
maintenance
1) The dust filter can be treated as household waste.
2) Refer to "Technical Data" section for the battery operating time.
134
Definition
Appropriate measures intended to retain the functional condition of a medi-
cal device
Measures intended to determine and assess the actual state of a medical
device
Repeated indicated measures intended to retain the functional condition of
a medical device
Measures intended to restore the functional condition of a medical device
after the failure of a device function
Frequency
Every two years
– Every two years
– When the battery no longer
remains charged for the speci-
fied operating time
Every two years
Every two years
Performed by
Service personnel
User
2)
.
Service personnel
Service personnel
Instructions for Use Oxylog 3000 plus SW 1.n