– Insp.term.%PIF
Inspiration termination criteria of pressure sup-
ported strokes, as percentage of the peak
inspiratory flow (PIF).
– Hose type
The selected hose type must match the hose
type in use. Otherwise a correct volume mea-
surement cannot be guaranteed.
– Cuvette type (optional)
NOTE
The cuvette windows of the reusable cuvette and
disposable cuvette have different optical proper-
ties. Therefore, the correct cuvette type must be
selected in the Setting menu. Otherwise the zero
point is shifted by up to ± 8 mmHg of CO
Instructions for Use Oxylog 3000 plus SW 1.n

Setting Pressure Support VC-SIMV/PS

The following can also be set on the display for
VC-SIMV:
– Setting on page 1: Pressure Support
above PEEP.
– Setting on page 2: When
0 mbar, the pressure rise time Slope can be
set.
Flat slope = long pressure rise time
Medium slope = medium pressure rise
time
Steep slope = short pressure rise time.
.
2
Operation
Δ
Psupp
Δ
Psupp is set above
69