Checking Readiness For Operation; Perform Device Check - Dräger Oxylog 3000 plus Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

Getting Started

Checking readiness for operation

The device check should be performed:
– Before every use of the device if the breathing
hose was changed.
– At least every six months.
The Oxylog 3000 plus interrupts the device check if
a fault is detected.
The relevant fault is indicated on the screen.
WARNING
The patient may be endangered if the device
check is not completed successfully.

Perform device check

The device check consists of the following steps:
– Connect the test lung
– Switch the device ON
– Check connections
– System check
– Power supply failure check.
The duration of the device check is approximately
3 minutes.
54
Connect the test lung
A
B
C
1 Make sure that the angled connector (A) is
connected to the flow sensor.
2 Connect the catheter mount (B) of the test lung,
diameter 7 mm, to the angled connector.
The catheter mount simulates the resistance of
the airways.
3 Connect the balloon (C) of the test lung. Refer
to "List of Accessories" on page 167.
NOTE
BTPS values of a test lung are not the same as the
BTPS values of a patient. The Oxylog 3000 plus
measures and adapts according to BTPS values of
a patient. Therefore, when a test lung is
connected, the MVe and VTe indicated on display
may differ from the MVe and VTe that is set by the
operator.